Activity
From 05/24/2018 to 06/22/2018
06/19/2018
- 05:02 PM Analyzing #198 (Закрыта): Determination of thickness inhomogeneity of thin silicon detector
06/18/2018
- 06:20 PM Analyzing #186: Number of 3He coincidences with 3H
Analysis of data with beryllium beam gives interesting results. Data
mbsdaq@nra161:/LynxOS/mbsusr/mbsdaq/mbsru...
06/16/2018
- 03:50 AM Analyzing #186: Number of 3He coincidences with 3H
We compared dE-E identification plots obtained from experimental data (run 14) with data used for calibrations giving...
06/15/2018
- 02:13 AM Analyzing #183: Express analysis of experimental data
Raw data may be converted into ROOT format in following way:
Go4 library must be available. One of the well w...
- 11:02 AM Analyzing #195 (Закрыта): First estimate of inhomogeneity of thin Si detector
Phenomena was studied deeply in issue 198.
06/14/2018
- 02:26 PM Analyzing #198: Determination of thickness inhomogeneity of thin silicon detector
the same macros was used for determination of thickness of 20um det, but this time each SQY_L strips was separated in...
06/07/2018
- 02:11 PM Analyzing #198: Determination of thickness inhomogeneity of thin silicon detector
I have contacted the Micron Semiconductors. There is no input control of the thickness uniformity. They rely only upo...
- 01:58 PM Analyzing #198: Determination of thickness inhomogeneity of thin silicon detector
files 159.93.80.161:/LynxOS/mbsusr/mbsdaq/mbsrun/exp201804/data/calib/si_after/si-1_si-20_35cm_0deg_new1_*.lmd were p...
- 11:48 AM Developing #187: Калибрация 1 мм кремневого детектора
For data obtained in measurements with SQL, by irradiating it from 35 cm with 90 and 45 angle. In this measureme...
06/06/2018
- 02:50 PM Developing #187: Калибрация 1 мм кремневого детектора
We should compare the count rate in peaks with what is expected. Nominal activity is 3*10^4 Bq. For a pixel size of...
- 12:08 PM Analyzing #198: Determination of thickness inhomogeneity of thin silicon detector
In data 159.93.80.161:/LynxOS/mbsusr/mbsdaq/mbsrun/exp201804/data/calib/si_after/si-1_si-20_35cm_0deg_new1_*.lmd ther...
06/05/2018
- 06:20 PM Analyzing #199: Determination of positions of silicone detectors
The script for determine position of detectors:
- 06:00 PM Analyzing #199: Determination of positions of silicone detectors
The active area of thin detector is 50mm:50mm and there are 16 stripes for axis X (from left to right SQY_R[0]->SQY_...
06/04/2018
- 12:36 PM Developing #184: Калибрация тонково кремневого детектора
From the data obtained in meauserements described in task 197 dead layer was calculated. calculations was made https:...
- 11:55 AM Analyzing #198: Determination of thickness inhomogeneity of thin silicon detector
In order to analyse the properties last strips (12-15) data from measurements, where alpha source was moved to the le...
06/01/2018
- 01:19 PM Analyzing #199: Determination of positions of silicone detectors
We used data file si-1_si-20_35cm_0deg_new1_0003.lmd We took into account that center of the thick 1mm detector is ...
- 10:37 AM Developing #197: Measurement of thickness of thin silicon detector
Several measurements were carried out. In all of them two detectors were used. 20 mcm (discription see in file 7306 r...
05/31/2018
- 10:19 AM Analyzing #199 (Открыта): Determination of positions of silicone detectors
Determine positions of silicon detectors used in post-production measurement of thin detector thickness.
We take i...
05/30/2018
- 04:56 PM Analyzing #198: Determination of thickness inhomogeneity of thin silicon detector
data from the files 159.93.80.161:/LynxOS/mbsusr/mbsdaq/mbsrun/exp201804/data/calib/si_after/si-1_si-20_35cm_45deg_ne...
- 02:10 PM Analyzing #198 (Закрыта): Determination of thickness inhomogeneity of thin silicon detector
Determinate the map of thickness of thin silicon detector by treatment of data acquired in task #197.
05/25/2018
- 12:26 PM Analyzing #195: First estimate of inhomogeneity of thin Si detector
Estimation of the inhomogeneity of thin Si detector was done by script. The input data is root file converted from si...
05/24/2018
- 05:21 PM Developing #187: Калибрация 1 мм кремневого детектора
For data obtained in measurements with SQL, by irradiating it from 35 cm with 90 angle. Also in this measurements amp...
- 05:07 PM Developing #197 (Открыта): Measurement of thickness of thin silicon detector
We need to determine thickness of the thin (assumed thickness ~20 mcm) silicon detector. It is very probable that we ...
- 04:38 PM Developing #196 (Открыта): Post-production measurements
There are some measurements of various detectors parameters which should be performed after main experimental run. Th...