Analyzing #195

Analyzing #183: Express analysis of experimental data

First estimate of inhomogeneity of thin Si detector

Added by Vratislav Chudoba over 6 years ago. Updated over 6 years ago.

Status:ЗакрытаStart date:05/18/2018
Priority:НормальныйDue date:
Assignee:Ivan Muzalevsky% Done:

100%

Category:Software
Target version:-

Description

First estimate of inhomogeneity of thin silicon detector may be based on treatment of data used for energy calibration (acquired with very low statistics) saved in file

159.93.80.161:/home/LynxOS/mbsusr/mbsdaq/mbsrun/exp201804/data/calib/si_20_03_*.lmd

where alpha source was oriented to the surface of the detector under the angle ~90° at distance approximately 40 cm.

New measurement with high statistics is needed. Available data will be reported as soon as possible.

ecalib_sq20_03_report.c Magnifier - script (2.25 KB) Roman Slepnev, 05/25/2018 12:25 PM

sq20_calib_03.txt Magnifier - data (698 Bytes) Roman Slepnev, 05/25/2018 12:25 PM

sq20_thickness.txt Magnifier - result (1.17 KB) Roman Slepnev, 05/25/2018 12:26 PM

History

#1 Updated by Vratislav Chudoba over 6 years ago

  • Description updated (diff)

#2 Updated by Roman Slepnev over 6 years ago

Estimation of the inhomogeneity of thin Si detector was done by script. The input data is root file converted from si_20_03_*.lmd. Then from this root file the file (sq20_calib_03.txt) with positions of the 7.63 alpha peak in 1mm SI in Left telescope after passing through the 20mkm of Si (SQ20) was obtained with analysis in ROOT. The procedure with fitting of the thickness of SQ20 was done with granulating of 1 mm detector by combining 4 X strips (32/4=8 ) and 16 Y strips. There is no data in Y strip 0.

#3 Updated by Roman Slepnev over 6 years ago

  • Assignee changed from Roman Slepnev to Ivan Muzalevsky

#4 Updated by Ivan Muzalevsky over 6 years ago

  • Status changed from Открыта to Закрыта
  • % Done changed from 0 to 100

Phenomena was studied deeply in issue 198.